Conductivity is a key property to check when selecting an electromagnetic flow meter, as the instrument needs to obtain a reliable measurement signal from the liquid. Describing the fluid simply as “water” is not enough. Treatment methods, temperature and mixtures within the process can change conductivity. Selection should therefore consider the lowest conductivity expected during operation together with the operating limit of the chosen model.
Even when conductivity is suitable, flow range, installation, material compatibility and electrical connections remain important for accurate measurement. This guide explains which information to prepare before requesting a quotation.

How does an electromagnetic flow meter work?
A conductive liquid moving through the magnetic field in the measuring tube generates an electrical voltage. The instrument relates this signal to flow velocity and uses the measuring cross-section to calculate volumetric flow. The result can be displayed in m³/h, for example.
With this method, the liquid’s electrical conductivity forms part of the measurement conditions. Insufficient conductivity for the instrument can cause an unstable signal or prevent measurement. However, higher conductivity does not always mean higher accuracy. Accuracy depends on the instrument’s specified operating range, application and installation.
Is there a single minimum conductivity value?
There is no single minimum conductivity value that applies to all electromagnetic flow meters. The lower limit can vary with sensor design, measuring diameter, electronics and, in some applications, connection conditions. It is not appropriate to apply another instrument’s limit directly to the model being selected.
Specify conductivity in µS/cm or mS/cm when requesting a quotation. Remember that 1 mS/cm = 1000 µS/cm. Also record the measurement temperature and whether the instrument uses temperature compensation. A laboratory result corrected to a reference temperature should not be interpreted in the same way as actual operating conditions in the line.
If conductivity is close to the model’s lower limit, do not rely on a single sample. Assess the lowest values expected during start-up, normal production, rinsing and product changes.
Which liquids need particular attention?
Process water and cooling water
Electromagnetic measurement can be considered for these lines, but the water’s chemical composition and additives should be known. Chemicals used in closed circuits, temperature and accumulated contamination can also affect material selection. Share minimum, normal and maximum flow rates rather than selecting a meter solely by existing pipe diameter.
Treated water and low-conductivity water
Processes such as reverse osmosis or deionisation can reduce conductivity. A statement that an instrument is “suitable for treated water” is not enough for selection. If the measured value does not meet the chosen model’s requirements, a different sensor design or measurement method should be considered.
Wastewater and liquids containing particles
Even with sufficient conductivity, solids, air bubbles, electrode deposits and abrasive particles require separate assessment. Electrode and lining materials must suit the fluid’s chemical and mechanical effects. Access for cleaning and maintenance should also be planned during selection.
Five installation checks beyond conductivity
- Measuring tube fullness: In standard full-pipe applications, the measuring cross-section must remain filled with liquid. Systems specially designed for partially filled pipes require separate assessment.
- Air and gas accumulation: Sensor location and flow conditions can affect electrode contact with the liquid and signal stability.
- Inlet and outlet conditions: Check the effects of elbows, valves, pumps and reducers. Determine the required straight pipe lengths from the chosen model’s installation requirements; do not apply the same distances to every instrument.
- Electrical reference and earthing: Use a connection arrangement suited to the pipe material and instrument design. Do not assume that the same earthing component is required in every application.
- Remote display and cabling: The cable, shielding and distance between sensor and electronics must meet the model’s requirements. These details are particularly important at low conductivity.
What information is needed for a technical quotation?
- Fluid name, composition and any additives or solids.
- Minimum, normal and maximum flow rates, with the flow units used.
- Measured conductivity range, measurement temperature and conditions.
- Operating and design pressure and temperature ranges.
- Pipe diameter, material, connection standard and available installation space.
- Position of pumps, valves and elbows relative to the sensor, and whether the pipe remains full.
- Power supply, display location and output signals required by the automation system.
For a broader preliminary assessment, see our guide to 10 flow meter selection checks as well.
Frequently asked questions
Can an electromagnetic flow meter measure steam?
This measurement method is intended for conductive liquids; it is not used to measure steam or gas flow. A steam line requires another method suited to its operating conditions.
Is any measuring diameter suitable if conductivity is sufficient?
No. Check the velocity at the chosen measuring diameter, performance at minimum flow and the permitted operating range. Pipe diameter alone does not determine flow meter diameter.
Does installing a flow meter directly save energy?
Measurement helps monitor consumption and investigate losses. Savings result from operational or equipment improvements based on the measurement data. A fixed saving cannot be attributed to installing an instrument alone.